Semi auto Chemistry analyzer

Description

Semi auto Chemistry analyzer

Chem – S1 Semi Automated

Biochemistry analyzer

SPECIFICATIONS:

Light Source: 6V 10W, halogen lamp of more than 2000 hours’ life time

Spectrometer: interference filter.

Wavelength: 340nm,380nm,405nm,505nm,546nm,578nm,620nm, and one wavelength free. CML precision: ±2nm

FWHM: ≤10nm

Stray light:≤0.5%(absorbance≥2.5) Absorbance range: -0.214~3.000 Absorbance linearity: ±1%

Absorbance repeatability: ±0.001

Absorbance resolution: 0.0001 Precision: ≤1.0%

Flow cell:permanent quartz flow cell, minimal volume is 32μL, and disposable flow cell is selectable

Sampling system: pressure auto releasing pumps ensure the accuracy of suction volume

Test Method: End point,Fixed time,Kinetic,Factor,Muli-standards,Bichromatic,Dual wavelengths Absorption volume: 200μL~3000μL

Sample volume: 10μL~50μL

Reagent: Open reagent

Display: LCD with high brightness

 

Test item
Liver ALT,ALP,GGT,TP,Alb,TBIL,DBIL
Heart LDH,CK,CK-MB,AST, a-HBDH
Lipids CHO,TG,HDL_C,LDL_C
Lipoprotein Apo A1,Apo B,Lp(a)
kidney BUN,CRE,UA
Ion Na,Mg,Ca,K,P,Cl
Other CO2-CP,AMS,LIP,GLU,IgA,IgG,IgM,C3,C4,CRP